Doppler Displacement Module -

                                               Model 200-203

* Measure wall thickening with a single epicardial transducer
* Multiple measurements with minimum trauma

The Doppler Displacement Module uses a single epicardial  transducer to monitor local
transmural wall function. This technique causes no trauma to the underlaying layers of the myocardium and as such this technique can 'map' local function across an ischemic area.

Displacement is monitored by transmitting gated bursts of ultrasound into the myocardium.
Structures in the myocardial muscle layers reflect the incident sound bursts. A range gate
can be adjusted to monitor the reflected signals from various transmural depths. As
structures pass through the range gate position the reflected echoes are shifted in frequency
by an amount proportional to their velocity.

The Model 203 circuits develop an audio signal, whose frequency is proportional to the
velocity of the moving layer. The velocity signal is integrated to generate a displacement
signal. The integration is performed on a 12-bit digital up-down counter.

Triton's unique design allows the Model 203 output to be calibrated in terms of incremental
displacement. Accurate internal calibrations for displacements of 1,2, 5, and 10mm are
developed from the internal quartz crystal clock and are switch selectable from the front
panel Run/Calibrate switch. This feature is not available on other displacement systems.

The Model 203 Module offers five switch-selectable means to reset its integrator. The
integrator can be reset either manually or at fixed time intervals. Alternately, it can be reset
by the transducer-sensed electrogram or by an external trigger signal which may or not be
synchronous with heart rate. No other system offers such a degree of user control over the
inherent limitations of an indefinite integral technique.

Epicardial Doppler Displacement Transducers

Single epicardial transducers used in conjunction with the 203 Displacement Module to
evaluate myocardial wall thickening fraction. Available with no connector (N), Dacron patch (P) or Cloverleaf wire
loops (C).

                                                           
Part No.                           Item                      
                                                             
DMT-10-1.0    Doppler Displacement Transducer                
              10 MHz, 1.0 mm circular crystal with           
              stainless steel wire leads. Specify patch      
              (P), wire loops (C) attachment or none (N).                                                                 

DMT-10-2.0    Doppler Displacement Transducer                
              10 MHz, 2.0 mm square crystal with stainless   
              steel wire leads. Specify patch (P), wire      
              loops (C) attachment or none (N).                                                                           

DMT-10-3.0    Doppler Displacement Transducer                
              10 MHz, 3.0 mm circular crystal with           
              stainless steel wire leads. Specify patch      
              (P), wire loops (C) attachment or none (N).                                                                 

DMT-10-4.0    Doppler Displacement Transducer                
              10 MHz, 4.0 mm square crystal with stainless   
              steel wire leads. Specify patch (P), wire      
              loops (C) attachment or none (N).              
                                                             
DMT-20-1.0    Doppler Displacement Transducer                
              20 MHz, 1.0 mm circular crystal with           
              stainless steel wire leads. Specify patch      
              (P), wire loops (C) attachment or none (N).    
                                                             
DMT-20-2.0    Doppler Displacement Transducer                
              20 MHz, 2.0 mm square crystal with stainless   
              steel wire leads. Specify patch (P), wire      
              loops (C) attachment or none (N).              
                                                             
DMT-20-3.0    Doppler Displacement Transducer                
              20 MHz, 3.0 mm circular crystal with           
              stainless steel wire leads. Specify patch      
              (P), wire loops (C) attachment or none (N).    
                                                             
DMT-20-4.0    Doppler Displacement Transducer                
              20 MHz, 4.0 mm square crystal with stainless   
              steel wire leads. Specify patch (P), wire      
              loops (C) attachment or none (N).             
                                                                                                                          