Pressure-dP/dt Amplifier Module -

                                            Model 200-205

* Measure blood pressure and compute dP/dt in real-time

Konigsberg and Extracorporeal Pressure Transducers

Konigsberg In-Vitro Pressure Transducer Calibration

The Pressure-dP/dt Amplifier Module is a precision bridge transducer amplifier and
regulated 5 volt DC bridge excitation source plus a differentiator and precision derivative
calibrator. The Model 205 will operate most implanted pressure transducers, pressure
catheters and extracorporeal gauges.

The Model 205 provides front panel adjustment for bridge balance. Calibration and zero are
generated with precision internal bridges and are selected with a front panel RUN/Calibrate
switch. The 'zero' output represents a balanced bridge and the calibrate position represents
100 mm Hg when used with a 5uV/V/mm Hg sensitivity gauge. The calibration output

can be used with other gauge sensitivities and the full-scale calibration value will be inversely
proportional to 5uV/V/mm Hg divided by the actual gauge factor.

The Model 205 provides both instantaneous and mean pressure output signals. The output
signals are controlled by a set of front panel gain and offset controls which make it possible
to configure the module to drive chart recorders, tape recorders or A/D converters.

In addition to the pressure outputs, the Pressure-dP/dt Module contains a precision low
noise differentiator with two-pole high frequency cut-off. The differentiator has independent
front panel gain and offset controls for use with various recording devices.

The Model 205 has a differentiating calibrator which is unique. Internal clock circuits
generate a multiple slope, constant amplitude waveform which is used to calibrate the dP/dt
output. A front panel selector switch sets calibration steps in units of the pressure gauge
calibration, either internal bridge or manometer calibration. The calibration units are in mm
Hg/sec. Independent front panel gain and offset controls set the dP/dt calibrator output level in order to match to the
pressure calibration signal.